Comparative studies of different vascular stents in terms of mechanical performances: Finite element analysis

Seung Kwan Cho, Won Pil Park, Eun Jung Cho, Dohyung Lim, Han Sung Kim, Jai Young Ko

Research output: Chapter in Book/Report/Conference proceedingConference contribution

3 Citations (Scopus)

Abstract

Seven representative commercialized stents were evaluated by FE analysis in terms of the criteria based on the itemized list of Food and Drug Administration (FDA) and European Standards (prEN). Tenax and MAC standard were generally excellent in various evaluation items. Palmaz-Schatz was the most excellent in fatigue durability, but its flexibility was a little inferior than that of other stents. This study may suggest a guideline for improvement in stent design and function.

Original languageEnglish
Title of host publication4th Latin American Congress on Biomedical Engineering 2007, Bioengineering Solutions for Latin America Health
PublisherSpringer Verlag
Pages744-747
Number of pages4
ISBN (Print)9783540744702
DOIs
Publication statusPublished - 2008
Event4th Latin American Congress on Biomedical Engineering, CLAIB 2007 - Margarita Island, Venezuela, Bolivarian Republic of
Duration: 2007 Sep 242007 Sep 28

Publication series

NameIFMBE Proceedings
Volume18
ISSN (Print)1680-0737

Other

Other4th Latin American Congress on Biomedical Engineering, CLAIB 2007
CountryVenezuela, Bolivarian Republic of
CityMargarita Island
Period07/9/2407/9/28

All Science Journal Classification (ASJC) codes

  • Bioengineering
  • Biomedical Engineering

Fingerprint Dive into the research topics of 'Comparative studies of different vascular stents in terms of mechanical performances: Finite element analysis'. Together they form a unique fingerprint.