Comparison of higher order aberrations in eyes with aspherical or spherical intraocular lenses

S. W. Kim, H. Ahn, E. K. Kim, T. I. Kim

Research output: Contribution to journalArticle

26 Citations (Scopus)

Abstract

Purpose To compare ocular, internal, corneal aberrations, and contrast sensitivity in patients implanted with aspheric or spherical intraocular lenses (IOLs).MethodsTotal, corneal, and internal aberrations were measured using ray-tracing technology in 23 eyes implanted with aspheric (Tecnis ZA9003) IOLs, and in 25 eyes implanted with spherical (AcrySof SA60AT) IOLs. Spherical aberration (SA), coma, trefoil, and RMS of higher order aberrations (HOAs) were analysed for 4.0 and 6.0 mm optical zones. Best-corrected visual acuities (BCVAs), contrast sensitivity under photopic and mesopic conditions, and modulation transfer function (MTF) were evaluated to compare visual performances.ResultsThe mean internal SA was 0.06/0.35 m in the aspheric IOL group and 0.04/0.17 m in the spherical IOL group (at 4 mm/6 mm pupil). Total SA was near to zero in the aspheric IOL group, and was 0.110.034/0.480.13 m in the spherical IOL group. The internal Z 3 1 and total Z 3 1 were more negative in the spherical IOL group. The internal Z 3 3 was more negative in the aspheric IOL group. The two groups were similar in terms of other Zernike values and RMS of HOA. There were no differences in BCVA between groups. The aspheric IOL group showed better contrast sensitivity at 3-6 cpd in photopic and 1.5-6 cpd in mesopic conditions. The MTF curve was better for the aspheric IOL group at 5-10 cpd.ConclusionsThe aspheric IOL with a negative SA reduced total SA by compensating for positive corneal SA. These effects resulted in improved contrast sensitivity.

Original languageEnglish
Pages (from-to)1493-1498
Number of pages6
JournalEye
Volume22
Issue number12
DOIs
Publication statusPublished - 2008 Dec

Fingerprint

Intraocular Lenses
Contrast Sensitivity
Visual Acuity
Pupil
Coma

All Science Journal Classification (ASJC) codes

  • Ophthalmology
  • Sensory Systems

Cite this

@article{8c73974abecb4d7f8317e43671278c00,
title = "Comparison of higher order aberrations in eyes with aspherical or spherical intraocular lenses",
abstract = "Purpose To compare ocular, internal, corneal aberrations, and contrast sensitivity in patients implanted with aspheric or spherical intraocular lenses (IOLs).MethodsTotal, corneal, and internal aberrations were measured using ray-tracing technology in 23 eyes implanted with aspheric (Tecnis ZA9003) IOLs, and in 25 eyes implanted with spherical (AcrySof SA60AT) IOLs. Spherical aberration (SA), coma, trefoil, and RMS of higher order aberrations (HOAs) were analysed for 4.0 and 6.0 mm optical zones. Best-corrected visual acuities (BCVAs), contrast sensitivity under photopic and mesopic conditions, and modulation transfer function (MTF) were evaluated to compare visual performances.ResultsThe mean internal SA was 0.06/0.35 m in the aspheric IOL group and 0.04/0.17 m in the spherical IOL group (at 4 mm/6 mm pupil). Total SA was near to zero in the aspheric IOL group, and was 0.110.034/0.480.13 m in the spherical IOL group. The internal Z 3 1 and total Z 3 1 were more negative in the spherical IOL group. The internal Z 3 3 was more negative in the aspheric IOL group. The two groups were similar in terms of other Zernike values and RMS of HOA. There were no differences in BCVA between groups. The aspheric IOL group showed better contrast sensitivity at 3-6 cpd in photopic and 1.5-6 cpd in mesopic conditions. The MTF curve was better for the aspheric IOL group at 5-10 cpd.ConclusionsThe aspheric IOL with a negative SA reduced total SA by compensating for positive corneal SA. These effects resulted in improved contrast sensitivity.",
author = "Kim, {S. W.} and H. Ahn and Kim, {E. K.} and Kim, {T. I.}",
year = "2008",
month = "12",
doi = "10.1038/eye.2008.302",
language = "English",
volume = "22",
pages = "1493--1498",
journal = "Eye",
issn = "0950-222X",
publisher = "Nature Publishing Group",
number = "12",

}

Comparison of higher order aberrations in eyes with aspherical or spherical intraocular lenses. / Kim, S. W.; Ahn, H.; Kim, E. K.; Kim, T. I.

In: Eye, Vol. 22, No. 12, 12.2008, p. 1493-1498.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Comparison of higher order aberrations in eyes with aspherical or spherical intraocular lenses

AU - Kim, S. W.

AU - Ahn, H.

AU - Kim, E. K.

AU - Kim, T. I.

PY - 2008/12

Y1 - 2008/12

N2 - Purpose To compare ocular, internal, corneal aberrations, and contrast sensitivity in patients implanted with aspheric or spherical intraocular lenses (IOLs).MethodsTotal, corneal, and internal aberrations were measured using ray-tracing technology in 23 eyes implanted with aspheric (Tecnis ZA9003) IOLs, and in 25 eyes implanted with spherical (AcrySof SA60AT) IOLs. Spherical aberration (SA), coma, trefoil, and RMS of higher order aberrations (HOAs) were analysed for 4.0 and 6.0 mm optical zones. Best-corrected visual acuities (BCVAs), contrast sensitivity under photopic and mesopic conditions, and modulation transfer function (MTF) were evaluated to compare visual performances.ResultsThe mean internal SA was 0.06/0.35 m in the aspheric IOL group and 0.04/0.17 m in the spherical IOL group (at 4 mm/6 mm pupil). Total SA was near to zero in the aspheric IOL group, and was 0.110.034/0.480.13 m in the spherical IOL group. The internal Z 3 1 and total Z 3 1 were more negative in the spherical IOL group. The internal Z 3 3 was more negative in the aspheric IOL group. The two groups were similar in terms of other Zernike values and RMS of HOA. There were no differences in BCVA between groups. The aspheric IOL group showed better contrast sensitivity at 3-6 cpd in photopic and 1.5-6 cpd in mesopic conditions. The MTF curve was better for the aspheric IOL group at 5-10 cpd.ConclusionsThe aspheric IOL with a negative SA reduced total SA by compensating for positive corneal SA. These effects resulted in improved contrast sensitivity.

AB - Purpose To compare ocular, internal, corneal aberrations, and contrast sensitivity in patients implanted with aspheric or spherical intraocular lenses (IOLs).MethodsTotal, corneal, and internal aberrations were measured using ray-tracing technology in 23 eyes implanted with aspheric (Tecnis ZA9003) IOLs, and in 25 eyes implanted with spherical (AcrySof SA60AT) IOLs. Spherical aberration (SA), coma, trefoil, and RMS of higher order aberrations (HOAs) were analysed for 4.0 and 6.0 mm optical zones. Best-corrected visual acuities (BCVAs), contrast sensitivity under photopic and mesopic conditions, and modulation transfer function (MTF) were evaluated to compare visual performances.ResultsThe mean internal SA was 0.06/0.35 m in the aspheric IOL group and 0.04/0.17 m in the spherical IOL group (at 4 mm/6 mm pupil). Total SA was near to zero in the aspheric IOL group, and was 0.110.034/0.480.13 m in the spherical IOL group. The internal Z 3 1 and total Z 3 1 were more negative in the spherical IOL group. The internal Z 3 3 was more negative in the aspheric IOL group. The two groups were similar in terms of other Zernike values and RMS of HOA. There were no differences in BCVA between groups. The aspheric IOL group showed better contrast sensitivity at 3-6 cpd in photopic and 1.5-6 cpd in mesopic conditions. The MTF curve was better for the aspheric IOL group at 5-10 cpd.ConclusionsThe aspheric IOL with a negative SA reduced total SA by compensating for positive corneal SA. These effects resulted in improved contrast sensitivity.

UR - http://www.scopus.com/inward/record.url?scp=65349090618&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=65349090618&partnerID=8YFLogxK

U2 - 10.1038/eye.2008.302

DO - 10.1038/eye.2008.302

M3 - Article

C2 - 18927594

AN - SCOPUS:65349090618

VL - 22

SP - 1493

EP - 1498

JO - Eye

JF - Eye

SN - 0950-222X

IS - 12

ER -