Design and performance evaluation of a laboratory-made 200 nm precut electrical cascade impactor

Jangseop Han, Joohee Seo, Dongho Park, Junho Hyun, Jungho Hwang

Research output: Contribution to journalArticle

Abstract

Increasing public concern regarding air quality has led to the development of efficient aerosol-monitoring techniques. Among the various aerosol measurement instruments based on electrical methods, in this study, an electrical cascade impactor (ECI) was designed and fabricated in our laboratory and was used to measure the real-time size distribution of submicron-sized aerosols. In the study by Park et al. (2007), it was assumed that the size distribution of incoming particles follows a unimodal lognormal distribution. However, in this study, the distribution of particles captured at each stage (including the Faraday cage) was assumed to be a unimodal lognormal distribution; hence, the incoming particles may follow any size distribution. After the particle charging characteristics were obtained for different particle sizes, experiments were performed with monodisperse test particles to determine the collection efficiency of each stage. The current measured in each stage was converted into a number based size distribution of aerosols by using the data inversion algorithm, which utilized the experimentally obtained collection efficiency. Then, a performance evaluation was performed, both in the laboratory and in the field. The results obtained by our ECI were in agreement with the scanning mobility particle sizer (SMPS) data.

Original languageEnglish
Pages (from-to)1118-1130
Number of pages13
JournalAerosol and Air Quality Research
Volume18
Issue number5
DOIs
Publication statusPublished - 2018 May 1

Fingerprint

Aerosols
aerosol
Air quality
electrical method
data inversion
Particle size
Scanning
impactor
particle
evaluation
laboratory
Monitoring
air quality
particle size
Experiments
distribution
experiment

All Science Journal Classification (ASJC) codes

  • Environmental Chemistry
  • Pollution

Cite this

Han, Jangseop ; Seo, Joohee ; Park, Dongho ; Hyun, Junho ; Hwang, Jungho. / Design and performance evaluation of a laboratory-made 200 nm precut electrical cascade impactor. In: Aerosol and Air Quality Research. 2018 ; Vol. 18, No. 5. pp. 1118-1130.
@article{e7026577a30e460f870279e64c091139,
title = "Design and performance evaluation of a laboratory-made 200 nm precut electrical cascade impactor",
abstract = "Increasing public concern regarding air quality has led to the development of efficient aerosol-monitoring techniques. Among the various aerosol measurement instruments based on electrical methods, in this study, an electrical cascade impactor (ECI) was designed and fabricated in our laboratory and was used to measure the real-time size distribution of submicron-sized aerosols. In the study by Park et al. (2007), it was assumed that the size distribution of incoming particles follows a unimodal lognormal distribution. However, in this study, the distribution of particles captured at each stage (including the Faraday cage) was assumed to be a unimodal lognormal distribution; hence, the incoming particles may follow any size distribution. After the particle charging characteristics were obtained for different particle sizes, experiments were performed with monodisperse test particles to determine the collection efficiency of each stage. The current measured in each stage was converted into a number based size distribution of aerosols by using the data inversion algorithm, which utilized the experimentally obtained collection efficiency. Then, a performance evaluation was performed, both in the laboratory and in the field. The results obtained by our ECI were in agreement with the scanning mobility particle sizer (SMPS) data.",
author = "Jangseop Han and Joohee Seo and Dongho Park and Junho Hyun and Jungho Hwang",
year = "2018",
month = "5",
day = "1",
doi = "10.4209/aaqr.2017.11.0471",
language = "English",
volume = "18",
pages = "1118--1130",
journal = "Aerosol and Air Quality Research",
issn = "1680-8584",
publisher = "AAGR Aerosol and Air Quality Research",
number = "5",

}

Design and performance evaluation of a laboratory-made 200 nm precut electrical cascade impactor. / Han, Jangseop; Seo, Joohee; Park, Dongho; Hyun, Junho; Hwang, Jungho.

In: Aerosol and Air Quality Research, Vol. 18, No. 5, 01.05.2018, p. 1118-1130.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Design and performance evaluation of a laboratory-made 200 nm precut electrical cascade impactor

AU - Han, Jangseop

AU - Seo, Joohee

AU - Park, Dongho

AU - Hyun, Junho

AU - Hwang, Jungho

PY - 2018/5/1

Y1 - 2018/5/1

N2 - Increasing public concern regarding air quality has led to the development of efficient aerosol-monitoring techniques. Among the various aerosol measurement instruments based on electrical methods, in this study, an electrical cascade impactor (ECI) was designed and fabricated in our laboratory and was used to measure the real-time size distribution of submicron-sized aerosols. In the study by Park et al. (2007), it was assumed that the size distribution of incoming particles follows a unimodal lognormal distribution. However, in this study, the distribution of particles captured at each stage (including the Faraday cage) was assumed to be a unimodal lognormal distribution; hence, the incoming particles may follow any size distribution. After the particle charging characteristics were obtained for different particle sizes, experiments were performed with monodisperse test particles to determine the collection efficiency of each stage. The current measured in each stage was converted into a number based size distribution of aerosols by using the data inversion algorithm, which utilized the experimentally obtained collection efficiency. Then, a performance evaluation was performed, both in the laboratory and in the field. The results obtained by our ECI were in agreement with the scanning mobility particle sizer (SMPS) data.

AB - Increasing public concern regarding air quality has led to the development of efficient aerosol-monitoring techniques. Among the various aerosol measurement instruments based on electrical methods, in this study, an electrical cascade impactor (ECI) was designed and fabricated in our laboratory and was used to measure the real-time size distribution of submicron-sized aerosols. In the study by Park et al. (2007), it was assumed that the size distribution of incoming particles follows a unimodal lognormal distribution. However, in this study, the distribution of particles captured at each stage (including the Faraday cage) was assumed to be a unimodal lognormal distribution; hence, the incoming particles may follow any size distribution. After the particle charging characteristics were obtained for different particle sizes, experiments were performed with monodisperse test particles to determine the collection efficiency of each stage. The current measured in each stage was converted into a number based size distribution of aerosols by using the data inversion algorithm, which utilized the experimentally obtained collection efficiency. Then, a performance evaluation was performed, both in the laboratory and in the field. The results obtained by our ECI were in agreement with the scanning mobility particle sizer (SMPS) data.

UR - http://www.scopus.com/inward/record.url?scp=85046772629&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85046772629&partnerID=8YFLogxK

U2 - 10.4209/aaqr.2017.11.0471

DO - 10.4209/aaqr.2017.11.0471

M3 - Article

AN - SCOPUS:85046772629

VL - 18

SP - 1118

EP - 1130

JO - Aerosol and Air Quality Research

JF - Aerosol and Air Quality Research

SN - 1680-8584

IS - 5

ER -