Development of a shear measurement sensor for measuring forces at human-machine interfaces

Young Kuen Cho, Seong Guk Kim, Donghyun Kim, Hyung Joo Kim, Jeicheong Ryu, Dohyung Lim, Chang Yong Ko, Han Sung Kim

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Measuring shear force is crucial for investigating the pathology and treatment of pressure ulcers. In this study, we introduced a bi-axial shear transducer based on strain gauges as a new shear sensor. The sensor consisted of aluminum and polyvinyl chloride plates placed between quadrangular aluminum plates. On the middle plate, two strain gauges were placed orthogonal to one another. The shear sensor (54mm×54mm×4.1mm), which was validated by using standard weights, displayed high accuracy and precision (measurement range, -50 to 50N; sensitivity, 0.3N; linear relationship, R2=0.9625; crosstalk error, 0.635%±0.031%; equipment variation, 4.183). The shear force on the interface between the human body and a stand-up wheelchair was measured during sitting or standing movements, using two mats (44.8cm×44.8cm per mat) that consisted of 24 shear sensors. Shear forces on the sacrum and ischium were almost five times higher (15.5N at last posture) than those on other sites (3.5N on average) during experiments periods. In conclusion, the proposed shear sensor may be reliable and useful for measuring the shear force on human-machine interfaces.

Original languageEnglish
Pages (from-to)1721-1728
Number of pages8
JournalMedical Engineering and Physics
Volume36
Issue number12
DOIs
Publication statusPublished - 2014 Jan 1

Fingerprint

Ischium
Sacrum
Wheelchairs
Pressure Ulcer
Aluminum
Transducers
Posture
Human Body
Polyvinyl Chloride
Pathology
Weights and Measures
Equipment and Supplies
Sensors
Strain gages
Aluminum chloride
Crosstalk
Polyvinyl chlorides
aluminum chloride

All Science Journal Classification (ASJC) codes

  • Biophysics
  • Biomedical Engineering

Cite this

Cho, Young Kuen ; Kim, Seong Guk ; Kim, Donghyun ; Kim, Hyung Joo ; Ryu, Jeicheong ; Lim, Dohyung ; Ko, Chang Yong ; Kim, Han Sung. / Development of a shear measurement sensor for measuring forces at human-machine interfaces. In: Medical Engineering and Physics. 2014 ; Vol. 36, No. 12. pp. 1721-1728.
@article{cf9c128057014f2482f18e1219b99f6e,
title = "Development of a shear measurement sensor for measuring forces at human-machine interfaces",
abstract = "Measuring shear force is crucial for investigating the pathology and treatment of pressure ulcers. In this study, we introduced a bi-axial shear transducer based on strain gauges as a new shear sensor. The sensor consisted of aluminum and polyvinyl chloride plates placed between quadrangular aluminum plates. On the middle plate, two strain gauges were placed orthogonal to one another. The shear sensor (54mm×54mm×4.1mm), which was validated by using standard weights, displayed high accuracy and precision (measurement range, -50 to 50N; sensitivity, 0.3N; linear relationship, R2=0.9625; crosstalk error, 0.635{\%}±0.031{\%}; equipment variation, 4.183). The shear force on the interface between the human body and a stand-up wheelchair was measured during sitting or standing movements, using two mats (44.8cm×44.8cm per mat) that consisted of 24 shear sensors. Shear forces on the sacrum and ischium were almost five times higher (15.5N at last posture) than those on other sites (3.5N on average) during experiments periods. In conclusion, the proposed shear sensor may be reliable and useful for measuring the shear force on human-machine interfaces.",
author = "Cho, {Young Kuen} and Kim, {Seong Guk} and Donghyun Kim and Kim, {Hyung Joo} and Jeicheong Ryu and Dohyung Lim and Ko, {Chang Yong} and Kim, {Han Sung}",
year = "2014",
month = "1",
day = "1",
doi = "10.1016/j.medengphy.2014.09.010",
language = "English",
volume = "36",
pages = "1721--1728",
journal = "Medical Engineering and Physics",
issn = "1350-4533",
publisher = "Elsevier BV",
number = "12",

}

Development of a shear measurement sensor for measuring forces at human-machine interfaces. / Cho, Young Kuen; Kim, Seong Guk; Kim, Donghyun; Kim, Hyung Joo; Ryu, Jeicheong; Lim, Dohyung; Ko, Chang Yong; Kim, Han Sung.

In: Medical Engineering and Physics, Vol. 36, No. 12, 01.01.2014, p. 1721-1728.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Development of a shear measurement sensor for measuring forces at human-machine interfaces

AU - Cho, Young Kuen

AU - Kim, Seong Guk

AU - Kim, Donghyun

AU - Kim, Hyung Joo

AU - Ryu, Jeicheong

AU - Lim, Dohyung

AU - Ko, Chang Yong

AU - Kim, Han Sung

PY - 2014/1/1

Y1 - 2014/1/1

N2 - Measuring shear force is crucial for investigating the pathology and treatment of pressure ulcers. In this study, we introduced a bi-axial shear transducer based on strain gauges as a new shear sensor. The sensor consisted of aluminum and polyvinyl chloride plates placed between quadrangular aluminum plates. On the middle plate, two strain gauges were placed orthogonal to one another. The shear sensor (54mm×54mm×4.1mm), which was validated by using standard weights, displayed high accuracy and precision (measurement range, -50 to 50N; sensitivity, 0.3N; linear relationship, R2=0.9625; crosstalk error, 0.635%±0.031%; equipment variation, 4.183). The shear force on the interface between the human body and a stand-up wheelchair was measured during sitting or standing movements, using two mats (44.8cm×44.8cm per mat) that consisted of 24 shear sensors. Shear forces on the sacrum and ischium were almost five times higher (15.5N at last posture) than those on other sites (3.5N on average) during experiments periods. In conclusion, the proposed shear sensor may be reliable and useful for measuring the shear force on human-machine interfaces.

AB - Measuring shear force is crucial for investigating the pathology and treatment of pressure ulcers. In this study, we introduced a bi-axial shear transducer based on strain gauges as a new shear sensor. The sensor consisted of aluminum and polyvinyl chloride plates placed between quadrangular aluminum plates. On the middle plate, two strain gauges were placed orthogonal to one another. The shear sensor (54mm×54mm×4.1mm), which was validated by using standard weights, displayed high accuracy and precision (measurement range, -50 to 50N; sensitivity, 0.3N; linear relationship, R2=0.9625; crosstalk error, 0.635%±0.031%; equipment variation, 4.183). The shear force on the interface between the human body and a stand-up wheelchair was measured during sitting or standing movements, using two mats (44.8cm×44.8cm per mat) that consisted of 24 shear sensors. Shear forces on the sacrum and ischium were almost five times higher (15.5N at last posture) than those on other sites (3.5N on average) during experiments periods. In conclusion, the proposed shear sensor may be reliable and useful for measuring the shear force on human-machine interfaces.

UR - http://www.scopus.com/inward/record.url?scp=84922799209&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84922799209&partnerID=8YFLogxK

U2 - 10.1016/j.medengphy.2014.09.010

DO - 10.1016/j.medengphy.2014.09.010

M3 - Article

C2 - 25445984

AN - SCOPUS:84922799209

VL - 36

SP - 1721

EP - 1728

JO - Medical Engineering and Physics

JF - Medical Engineering and Physics

SN - 1350-4533

IS - 12

ER -