Evaluation of ultrasound tissue damage based on changes in image echogenicity in canine kidney

Jongbum Seo, Timothy L. Hall, Matthew O'Donnell, Binh C. Tran, J. Brian Fowlkes, Charles A. Cain, Gerald D. Abrams

Research output: Contribution to journalArticle

11 Citations (Scopus)

Abstract

Sufficiently high intensity ultrasound can create hyperechoic regions in an ultrasound image due to local bubble generation. We explore the link between the temporal extent of these hyperechoic regions and tissue damage caused by ultrasound therapy. The decay rate of increased echogenicity from the focal zone in insonated live exteriorized canine kidney was quantified and correlated to the spatial extent of tissue damage. The decay half-time, t half , defined as the time for echogenicity enhancement to decay by a factor of 2, was observed in all cases to be greater than 41 s in spatial zones in which extensive histological damage was observed. In cases in which the measured t half was less than 11 s, the damage was limited to minor hemorrhage, or it was not detected. These t half discrimination boundaries of 41 and 11 s were not statistically different for cases in which contrast agent was used to enhance therapeutic efficiency. This was true even though contrast agent infusion significantly reduced the therapy pulse duration threshold for damage production.

Original languageEnglish
Pages (from-to)1111-1119
Number of pages9
JournalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Volume52
Issue number7
DOIs
Publication statusPublished - 2005 Jul 1

Fingerprint

kidneys
Ultrasonics
Tissue
damage
evaluation
therapy
hemorrhages
decay
decay rates
discrimination
pulse duration
bubbles
thresholds
augmentation

All Science Journal Classification (ASJC) codes

  • Instrumentation
  • Acoustics and Ultrasonics
  • Electrical and Electronic Engineering

Cite this

Seo, Jongbum ; Hall, Timothy L. ; O'Donnell, Matthew ; Tran, Binh C. ; Fowlkes, J. Brian ; Cain, Charles A. ; Abrams, Gerald D. / Evaluation of ultrasound tissue damage based on changes in image echogenicity in canine kidney. In: IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control. 2005 ; Vol. 52, No. 7. pp. 1111-1119.
@article{a4296939b7864a9cafe50649d32a7b41,
title = "Evaluation of ultrasound tissue damage based on changes in image echogenicity in canine kidney",
abstract = "Sufficiently high intensity ultrasound can create hyperechoic regions in an ultrasound image due to local bubble generation. We explore the link between the temporal extent of these hyperechoic regions and tissue damage caused by ultrasound therapy. The decay rate of increased echogenicity from the focal zone in insonated live exteriorized canine kidney was quantified and correlated to the spatial extent of tissue damage. The decay half-time, t half , defined as the time for echogenicity enhancement to decay by a factor of 2, was observed in all cases to be greater than 41 s in spatial zones in which extensive histological damage was observed. In cases in which the measured t half was less than 11 s, the damage was limited to minor hemorrhage, or it was not detected. These t half discrimination boundaries of 41 and 11 s were not statistically different for cases in which contrast agent was used to enhance therapeutic efficiency. This was true even though contrast agent infusion significantly reduced the therapy pulse duration threshold for damage production.",
author = "Jongbum Seo and Hall, {Timothy L.} and Matthew O'Donnell and Tran, {Binh C.} and Fowlkes, {J. Brian} and Cain, {Charles A.} and Abrams, {Gerald D.}",
year = "2005",
month = "7",
day = "1",
doi = "10.1109/TUFFC.2005.1503997",
language = "English",
volume = "52",
pages = "1111--1119",
journal = "IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control",
issn = "0885-3010",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "7",

}

Evaluation of ultrasound tissue damage based on changes in image echogenicity in canine kidney. / Seo, Jongbum; Hall, Timothy L.; O'Donnell, Matthew; Tran, Binh C.; Fowlkes, J. Brian; Cain, Charles A.; Abrams, Gerald D.

In: IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, Vol. 52, No. 7, 01.07.2005, p. 1111-1119.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Evaluation of ultrasound tissue damage based on changes in image echogenicity in canine kidney

AU - Seo, Jongbum

AU - Hall, Timothy L.

AU - O'Donnell, Matthew

AU - Tran, Binh C.

AU - Fowlkes, J. Brian

AU - Cain, Charles A.

AU - Abrams, Gerald D.

PY - 2005/7/1

Y1 - 2005/7/1

N2 - Sufficiently high intensity ultrasound can create hyperechoic regions in an ultrasound image due to local bubble generation. We explore the link between the temporal extent of these hyperechoic regions and tissue damage caused by ultrasound therapy. The decay rate of increased echogenicity from the focal zone in insonated live exteriorized canine kidney was quantified and correlated to the spatial extent of tissue damage. The decay half-time, t half , defined as the time for echogenicity enhancement to decay by a factor of 2, was observed in all cases to be greater than 41 s in spatial zones in which extensive histological damage was observed. In cases in which the measured t half was less than 11 s, the damage was limited to minor hemorrhage, or it was not detected. These t half discrimination boundaries of 41 and 11 s were not statistically different for cases in which contrast agent was used to enhance therapeutic efficiency. This was true even though contrast agent infusion significantly reduced the therapy pulse duration threshold for damage production.

AB - Sufficiently high intensity ultrasound can create hyperechoic regions in an ultrasound image due to local bubble generation. We explore the link between the temporal extent of these hyperechoic regions and tissue damage caused by ultrasound therapy. The decay rate of increased echogenicity from the focal zone in insonated live exteriorized canine kidney was quantified and correlated to the spatial extent of tissue damage. The decay half-time, t half , defined as the time for echogenicity enhancement to decay by a factor of 2, was observed in all cases to be greater than 41 s in spatial zones in which extensive histological damage was observed. In cases in which the measured t half was less than 11 s, the damage was limited to minor hemorrhage, or it was not detected. These t half discrimination boundaries of 41 and 11 s were not statistically different for cases in which contrast agent was used to enhance therapeutic efficiency. This was true even though contrast agent infusion significantly reduced the therapy pulse duration threshold for damage production.

UR - http://www.scopus.com/inward/record.url?scp=24944447993&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=24944447993&partnerID=8YFLogxK

U2 - 10.1109/TUFFC.2005.1503997

DO - 10.1109/TUFFC.2005.1503997

M3 - Article

C2 - 16212250

AN - SCOPUS:24944447993

VL - 52

SP - 1111

EP - 1119

JO - IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control

JF - IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control

SN - 0885-3010

IS - 7

ER -