Abstract

Background: By conventional methods, the identification of anaerobic bacteria is more time consuming and requires more expertise than the identification of aerobic bacteria. Although the matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) systems are relatively less studied, they have been reported to be a promising method for the identification of anaerobes. We evaluated the performance of the VITEK MS in vitro diagnostic (IVD; 1.1 database; bioMérieux, France) in the identification of anaerobes.

Methods: We used 274 anaerobic bacteria isolated from various clinical specimens. The results for the identification of the bacteria by VITEK MS were compared to those obtained by phenotypic methods and 16S rRNA gene sequencing.

Results: Among the 249 isolates included in the IVD database, the VITEK MS correctly identified 209 (83.9%) isolates to the species level and an additional 18 (7.2%) at the genus level. In particular, the VITEK MS correctly identified clinically relevant and frequently isolated anaerobic bacteria to the species level. The remaining 22 isolates (8.8%) were either not identified or misidentified. The VITEK MS could not identify the 25 isolates absent from the IVD database to the species level.

Conclusions: The VITEK MS showed reliable identifications for clinically relevant anaerobic bacteria.

Original languageEnglish
Pages (from-to)69-75
Number of pages7
JournalAnnals of Laboratory Medicine
Volume35
Issue number1
DOIs
Publication statusPublished - 2015 Jan 1

Fingerprint

Anaerobic Bacteria
Matrix-Assisted Laser Desorption-Ionization Mass Spectrometry
Ionization
Mass spectrometry
Desorption
Mass Spectrometry
Bacteria
Lasers
Databases
Aerobic bacteria
Aerobic Bacteria
rRNA Genes
France
Genes

All Science Journal Classification (ASJC) codes

  • Clinical Biochemistry
  • Biochemistry, medical

Cite this

@article{9e45903102954653b32132a39493b6f3,
title = "Evaluation of VITEK mass spectrometry (MS), a matrix-assisted laser desorption ionization time-of-flight MS system for identification of anaerobic bacteria",
abstract = "Background: By conventional methods, the identification of anaerobic bacteria is more time consuming and requires more expertise than the identification of aerobic bacteria. Although the matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) systems are relatively less studied, they have been reported to be a promising method for the identification of anaerobes. We evaluated the performance of the VITEK MS in vitro diagnostic (IVD; 1.1 database; bioM{\'e}rieux, France) in the identification of anaerobes.Methods: We used 274 anaerobic bacteria isolated from various clinical specimens. The results for the identification of the bacteria by VITEK MS were compared to those obtained by phenotypic methods and 16S rRNA gene sequencing.Results: Among the 249 isolates included in the IVD database, the VITEK MS correctly identified 209 (83.9{\%}) isolates to the species level and an additional 18 (7.2{\%}) at the genus level. In particular, the VITEK MS correctly identified clinically relevant and frequently isolated anaerobic bacteria to the species level. The remaining 22 isolates (8.8{\%}) were either not identified or misidentified. The VITEK MS could not identify the 25 isolates absent from the IVD database to the species level.Conclusions: The VITEK MS showed reliable identifications for clinically relevant anaerobic bacteria.",
author = "Wonmok Lee and Myungsook Kim and DongEun Yong and Seokhoon Jeong and Kyungwon Lee and Yunsop Chong",
year = "2015",
month = "1",
day = "1",
doi = "10.3343/alm.2015.35.1.69",
language = "English",
volume = "35",
pages = "69--75",
journal = "Annals of Laboratory Medicine",
issn = "2234-3806",
publisher = "Seoul National University",
number = "1",

}

TY - JOUR

T1 - Evaluation of VITEK mass spectrometry (MS), a matrix-assisted laser desorption ionization time-of-flight MS system for identification of anaerobic bacteria

AU - Lee, Wonmok

AU - Kim, Myungsook

AU - Yong, DongEun

AU - Jeong, Seokhoon

AU - Lee, Kyungwon

AU - Chong, Yunsop

PY - 2015/1/1

Y1 - 2015/1/1

N2 - Background: By conventional methods, the identification of anaerobic bacteria is more time consuming and requires more expertise than the identification of aerobic bacteria. Although the matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) systems are relatively less studied, they have been reported to be a promising method for the identification of anaerobes. We evaluated the performance of the VITEK MS in vitro diagnostic (IVD; 1.1 database; bioMérieux, France) in the identification of anaerobes.Methods: We used 274 anaerobic bacteria isolated from various clinical specimens. The results for the identification of the bacteria by VITEK MS were compared to those obtained by phenotypic methods and 16S rRNA gene sequencing.Results: Among the 249 isolates included in the IVD database, the VITEK MS correctly identified 209 (83.9%) isolates to the species level and an additional 18 (7.2%) at the genus level. In particular, the VITEK MS correctly identified clinically relevant and frequently isolated anaerobic bacteria to the species level. The remaining 22 isolates (8.8%) were either not identified or misidentified. The VITEK MS could not identify the 25 isolates absent from the IVD database to the species level.Conclusions: The VITEK MS showed reliable identifications for clinically relevant anaerobic bacteria.

AB - Background: By conventional methods, the identification of anaerobic bacteria is more time consuming and requires more expertise than the identification of aerobic bacteria. Although the matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) systems are relatively less studied, they have been reported to be a promising method for the identification of anaerobes. We evaluated the performance of the VITEK MS in vitro diagnostic (IVD; 1.1 database; bioMérieux, France) in the identification of anaerobes.Methods: We used 274 anaerobic bacteria isolated from various clinical specimens. The results for the identification of the bacteria by VITEK MS were compared to those obtained by phenotypic methods and 16S rRNA gene sequencing.Results: Among the 249 isolates included in the IVD database, the VITEK MS correctly identified 209 (83.9%) isolates to the species level and an additional 18 (7.2%) at the genus level. In particular, the VITEK MS correctly identified clinically relevant and frequently isolated anaerobic bacteria to the species level. The remaining 22 isolates (8.8%) were either not identified or misidentified. The VITEK MS could not identify the 25 isolates absent from the IVD database to the species level.Conclusions: The VITEK MS showed reliable identifications for clinically relevant anaerobic bacteria.

UR - http://www.scopus.com/inward/record.url?scp=84920508813&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84920508813&partnerID=8YFLogxK

U2 - 10.3343/alm.2015.35.1.69

DO - 10.3343/alm.2015.35.1.69

M3 - Article

C2 - 25553283

AN - SCOPUS:84920508813

VL - 35

SP - 69

EP - 75

JO - Annals of Laboratory Medicine

JF - Annals of Laboratory Medicine

SN - 2234-3806

IS - 1