Measurement of sensitivity level pressure corrections for LS2P laboratory standard microphones

Lixue Wu, George S.K. Wong, Peter Hanes, Won Suk Ohm

Research output: Contribution to journalArticlepeer-review

9 Citations (Scopus)

Abstract

The corrections for the influence of atmospheric pressure on the sensitivity levels of laboratory standard microphones vary with frequency. The sensitivity level pressure corrections for Brüel and Kjær 4180 and G.R.A.S. 40AG laboratory standard microphones were measured. Over the barometric pressure range from 94kPa to 106kPa, the sensitivity levels of the model 4180 and model 40AG laboratory standard microphones changed by 0.05 dB at 1 kHz and 0.12 dB at 20kHz, respectively. Empirical equations for sensitivity level pressure corrections for the above microphones are presented.

Original languageEnglish
Pages (from-to)45-48
Number of pages4
JournalMetrologia
Volume42
Issue number1
DOIs
Publication statusPublished - 2005 Feb

All Science Journal Classification (ASJC) codes

  • Engineering(all)

Fingerprint

Dive into the research topics of 'Measurement of sensitivity level pressure corrections for LS2P laboratory standard microphones'. Together they form a unique fingerprint.