Performance Evaluation of a New Automated Chemiluminescent Immunoanalyzer-Based Interferon-Gamma Releasing Assay AdvanSure I3 in Comparison With the QuantiFERON-TB Gold In-Tube Assay

Jin Ju Kim, Younhee Park, Dasom Choi, Hyon Suk Kim

Research output: Contribution to journalArticle

Abstract

BACKGROUND: The interferon-gamma (IFN-γ) releasing assay (IGRA) is widely used for latent tuberculosis infection (LTBI) diagnosis. We evaluated the analytical performance of a new automated chemiluminescent immunoanalyzer-based IGRA (CLIA-IGRA), AdvanSure I3 (LG Life Sciences, Seoul, Korea) and compared it with that of the QuantiFERON-TB Gold In-Tube (QFT-GIT) assay. METHODS: Repeatability and reproducibility were evaluated at four levels. Detection capability, including limit of blank (LoB), limit of detection (LoD), and limit of quantification (LoQ), was evaluated using IFN-γ standard material (National Institute for Biological Standards and Control code: 87/586). Agreement between the results of two assays was evaluated using 341 blood samples from healthcare workers and patients at a tertiary care hospital. To determine the cut-off value of CLIA-IGRA for diagnosing LTBI, the ROC curve was analyzed. RESULTS: Repeatability and reproducibility were 4.86-7.00% and 6.36-7.88% CV, respectively. LoB, LoD, and LoQ were 0.022, 0.077, and 0.249 IU/mL, respectively. IFN-γ values between CLIA-IGRA and QFT-GIT showed a strong correlation within the analytical measurable range of both assays, especially when the value was low. Qualitative comparison of the two assays yielded a 99.1% overall agreement (kappa coefficient=0.98). A cut-off value of 0.35 IU/mL was appropriate for diagnosing LTBI. CONCLUSIONS: CLIA-IGRA is a reliable assay for LTBI diagnosis, with performance similar to that of QFT-GIT.

Original languageEnglish
Pages (from-to)33-39
Number of pages7
JournalAnnals of laboratory medicine
Volume40
Issue number1
DOIs
Publication statusPublished - 2020 Jan 1

Fingerprint

Latent Tuberculosis
Gold
Interferon-gamma
Assays
Limit of Detection
Biological Science Disciplines
Tertiary Healthcare
Korea
Tertiary Care Centers
ROC Curve
Medicine
Delivery of Health Care
Blood

All Science Journal Classification (ASJC) codes

  • Clinical Biochemistry
  • Biochemistry, medical

Cite this

@article{964005b66944451f87d2f62126233e92,
title = "Performance Evaluation of a New Automated Chemiluminescent Immunoanalyzer-Based Interferon-Gamma Releasing Assay AdvanSure I3 in Comparison With the QuantiFERON-TB Gold In-Tube Assay",
abstract = "BACKGROUND: The interferon-gamma (IFN-γ) releasing assay (IGRA) is widely used for latent tuberculosis infection (LTBI) diagnosis. We evaluated the analytical performance of a new automated chemiluminescent immunoanalyzer-based IGRA (CLIA-IGRA), AdvanSure I3 (LG Life Sciences, Seoul, Korea) and compared it with that of the QuantiFERON-TB Gold In-Tube (QFT-GIT) assay. METHODS: Repeatability and reproducibility were evaluated at four levels. Detection capability, including limit of blank (LoB), limit of detection (LoD), and limit of quantification (LoQ), was evaluated using IFN-γ standard material (National Institute for Biological Standards and Control code: 87/586). Agreement between the results of two assays was evaluated using 341 blood samples from healthcare workers and patients at a tertiary care hospital. To determine the cut-off value of CLIA-IGRA for diagnosing LTBI, the ROC curve was analyzed. RESULTS: Repeatability and reproducibility were 4.86-7.00{\%} and 6.36-7.88{\%} CV, respectively. LoB, LoD, and LoQ were 0.022, 0.077, and 0.249 IU/mL, respectively. IFN-γ values between CLIA-IGRA and QFT-GIT showed a strong correlation within the analytical measurable range of both assays, especially when the value was low. Qualitative comparison of the two assays yielded a 99.1{\%} overall agreement (kappa coefficient=0.98). A cut-off value of 0.35 IU/mL was appropriate for diagnosing LTBI. CONCLUSIONS: CLIA-IGRA is a reliable assay for LTBI diagnosis, with performance similar to that of QFT-GIT.",
author = "Kim, {Jin Ju} and Younhee Park and Dasom Choi and Kim, {Hyon Suk}",
year = "2020",
month = "1",
day = "1",
doi = "10.3343/alm.2020.40.1.33",
language = "English",
volume = "40",
pages = "33--39",
journal = "Annals of Laboratory Medicine",
issn = "2234-3806",
publisher = "Seoul National University",
number = "1",

}

TY - JOUR

T1 - Performance Evaluation of a New Automated Chemiluminescent Immunoanalyzer-Based Interferon-Gamma Releasing Assay AdvanSure I3 in Comparison With the QuantiFERON-TB Gold In-Tube Assay

AU - Kim, Jin Ju

AU - Park, Younhee

AU - Choi, Dasom

AU - Kim, Hyon Suk

PY - 2020/1/1

Y1 - 2020/1/1

N2 - BACKGROUND: The interferon-gamma (IFN-γ) releasing assay (IGRA) is widely used for latent tuberculosis infection (LTBI) diagnosis. We evaluated the analytical performance of a new automated chemiluminescent immunoanalyzer-based IGRA (CLIA-IGRA), AdvanSure I3 (LG Life Sciences, Seoul, Korea) and compared it with that of the QuantiFERON-TB Gold In-Tube (QFT-GIT) assay. METHODS: Repeatability and reproducibility were evaluated at four levels. Detection capability, including limit of blank (LoB), limit of detection (LoD), and limit of quantification (LoQ), was evaluated using IFN-γ standard material (National Institute for Biological Standards and Control code: 87/586). Agreement between the results of two assays was evaluated using 341 blood samples from healthcare workers and patients at a tertiary care hospital. To determine the cut-off value of CLIA-IGRA for diagnosing LTBI, the ROC curve was analyzed. RESULTS: Repeatability and reproducibility were 4.86-7.00% and 6.36-7.88% CV, respectively. LoB, LoD, and LoQ were 0.022, 0.077, and 0.249 IU/mL, respectively. IFN-γ values between CLIA-IGRA and QFT-GIT showed a strong correlation within the analytical measurable range of both assays, especially when the value was low. Qualitative comparison of the two assays yielded a 99.1% overall agreement (kappa coefficient=0.98). A cut-off value of 0.35 IU/mL was appropriate for diagnosing LTBI. CONCLUSIONS: CLIA-IGRA is a reliable assay for LTBI diagnosis, with performance similar to that of QFT-GIT.

AB - BACKGROUND: The interferon-gamma (IFN-γ) releasing assay (IGRA) is widely used for latent tuberculosis infection (LTBI) diagnosis. We evaluated the analytical performance of a new automated chemiluminescent immunoanalyzer-based IGRA (CLIA-IGRA), AdvanSure I3 (LG Life Sciences, Seoul, Korea) and compared it with that of the QuantiFERON-TB Gold In-Tube (QFT-GIT) assay. METHODS: Repeatability and reproducibility were evaluated at four levels. Detection capability, including limit of blank (LoB), limit of detection (LoD), and limit of quantification (LoQ), was evaluated using IFN-γ standard material (National Institute for Biological Standards and Control code: 87/586). Agreement between the results of two assays was evaluated using 341 blood samples from healthcare workers and patients at a tertiary care hospital. To determine the cut-off value of CLIA-IGRA for diagnosing LTBI, the ROC curve was analyzed. RESULTS: Repeatability and reproducibility were 4.86-7.00% and 6.36-7.88% CV, respectively. LoB, LoD, and LoQ were 0.022, 0.077, and 0.249 IU/mL, respectively. IFN-γ values between CLIA-IGRA and QFT-GIT showed a strong correlation within the analytical measurable range of both assays, especially when the value was low. Qualitative comparison of the two assays yielded a 99.1% overall agreement (kappa coefficient=0.98). A cut-off value of 0.35 IU/mL was appropriate for diagnosing LTBI. CONCLUSIONS: CLIA-IGRA is a reliable assay for LTBI diagnosis, with performance similar to that of QFT-GIT.

UR - http://www.scopus.com/inward/record.url?scp=85071774318&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85071774318&partnerID=8YFLogxK

U2 - 10.3343/alm.2020.40.1.33

DO - 10.3343/alm.2020.40.1.33

M3 - Article

C2 - 31432637

AN - SCOPUS:85071774318

VL - 40

SP - 33

EP - 39

JO - Annals of Laboratory Medicine

JF - Annals of Laboratory Medicine

SN - 2234-3806

IS - 1

ER -