Abstract

Hepatitis B virus (HBV) nucleic acid amplification testing (NAAT) is important for the diagnosis and management of HBV infection. We evaluated the analytical performance of the cobas HBV NAAT (Roche Diagnostics GmbH, Mannheim, Germany) on the cobas 4800 System in comparison with COBAS AmpliPrep/COBAS TaqMan HBV Test (CAP/CTM HBV). Precision was evaluated using three levels of cobas HBV/HCV/HIV-1 Control Kit, and linearity was evaluated across the anticipated measuring range (10.0-1.0×10 9 IU/mL) at seven levels using clinical samples. Detection capability, including limit of blank (LOB), limit of detection (LOD) and limit of quantitation (LOQ), was verified using the 4th WHO International Standard for HBV DNA for NAT (NIBSC code: 10/266). Correlation between the two systems was compared using 205 clinical samples (102 sera and 103 EDTA plasma). Repeatability and total imprecision (coefficient of variation) ranged from 0.5% to 3.8% and from 0.5% to 3.5%, respectively. Linearity (coefficient of determination, R 2 ) was 0.999. LOB, LOD and LOQ were all acceptable within the observed proportion rate (85%). Correlation was very high between the two systems in both serum and plasma samples (correlation coefficient [r]=0.995). The new cobas HBV real-time PCR assay on the cobas 4800 System showed reliable analytical performances.

Original languageEnglish
Pages (from-to)1133-1139
Number of pages7
JournalClinical Chemistry and Laboratory Medicine
Volume56
Issue number7
DOIs
Publication statusPublished - 2018 Jun 27

Bibliographical note

Funding Information:
Acknowledgments: This study was supported by the grant and product from Roche Diagnostics. Author contributions: Kim H performed the experiment, analyzed the data and wrote the draft; Hur M designed the study and finalized the draft; Bae E, Lee KA and Lee WI participated in sample collection, reviewed the draft and commented on it. All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission. Research funding: None declared. Employment or leadership: None declared. Honorarium: None declared. Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

All Science Journal Classification (ASJC) codes

  • Clinical Biochemistry
  • Biochemistry, medical

Fingerprint Dive into the research topics of 'Performance evaluation of cobas HBV real-time PCR assay on Roche cobas 4800 System in comparison with COBAS AmpliPrep/COBAS TaqMan HBV Test'. Together they form a unique fingerprint.

Cite this