Spontaneous fracture of an implanted posterior chamber polyimide intraocular lens haptic: A case report

Haemin Kang, Kyung Eun Han, Tae Im Kim, Eung Kweon Kim

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

A 57-year-old male patient visited our clinic for decreased visual acuity in the right eye for 10 days. He denied any trauma history, but recalled that the symptom developed after straining. He had undergone uncomplicated phacoemulsification and posterior chamber intraocular lens (IOL) implantation in the bag of the right eye 11 years ago. The IOL was a three-piece silicone polyimide-haptics design. On slit-lamp examination, the IOL optic and proximal part of nasal fractured haptic were found in the anterior chamber. The distal part of fractured haptic was observed in the capsular bag. He underwent IOL exchange. The fracture site of the haptic was near the optic-haptic junction. This is the unique case report of a spontaneous fracture of an implanted posterior chamber polyimide IOL haptic, which implies the possibility of IOL haptic fracture in various haptic materials.

Original languageEnglish
Pages (from-to)348-350
Number of pages3
JournalIndian Journal of Ophthalmology
Volume62
Issue number3
DOIs
Publication statusPublished - 2014 Mar

Fingerprint

Spontaneous Fractures
Intraocular Lenses
Intraocular Lens Implantation
Phacoemulsification
Anterior Chamber
Silicones
Nose
Visual Acuity
Wounds and Injuries

All Science Journal Classification (ASJC) codes

  • Ophthalmology
  • Medicine(all)

Cite this

@article{361d6d2978ee4bb58cb425f32c5cc599,
title = "Spontaneous fracture of an implanted posterior chamber polyimide intraocular lens haptic: A case report",
abstract = "A 57-year-old male patient visited our clinic for decreased visual acuity in the right eye for 10 days. He denied any trauma history, but recalled that the symptom developed after straining. He had undergone uncomplicated phacoemulsification and posterior chamber intraocular lens (IOL) implantation in the bag of the right eye 11 years ago. The IOL was a three-piece silicone polyimide-haptics design. On slit-lamp examination, the IOL optic and proximal part of nasal fractured haptic were found in the anterior chamber. The distal part of fractured haptic was observed in the capsular bag. He underwent IOL exchange. The fracture site of the haptic was near the optic-haptic junction. This is the unique case report of a spontaneous fracture of an implanted posterior chamber polyimide IOL haptic, which implies the possibility of IOL haptic fracture in various haptic materials.",
author = "Haemin Kang and Han, {Kyung Eun} and Kim, {Tae Im} and Kim, {Eung Kweon}",
year = "2014",
month = "3",
doi = "10.4103/0301-4738.111195",
language = "English",
volume = "62",
pages = "348--350",
journal = "Indian Journal of Ophthalmology",
issn = "0301-4738",
publisher = "Medknow Publications and Media Pvt. Ltd",
number = "3",

}

Spontaneous fracture of an implanted posterior chamber polyimide intraocular lens haptic : A case report. / Kang, Haemin; Han, Kyung Eun; Kim, Tae Im; Kim, Eung Kweon.

In: Indian Journal of Ophthalmology, Vol. 62, No. 3, 03.2014, p. 348-350.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Spontaneous fracture of an implanted posterior chamber polyimide intraocular lens haptic

T2 - A case report

AU - Kang, Haemin

AU - Han, Kyung Eun

AU - Kim, Tae Im

AU - Kim, Eung Kweon

PY - 2014/3

Y1 - 2014/3

N2 - A 57-year-old male patient visited our clinic for decreased visual acuity in the right eye for 10 days. He denied any trauma history, but recalled that the symptom developed after straining. He had undergone uncomplicated phacoemulsification and posterior chamber intraocular lens (IOL) implantation in the bag of the right eye 11 years ago. The IOL was a three-piece silicone polyimide-haptics design. On slit-lamp examination, the IOL optic and proximal part of nasal fractured haptic were found in the anterior chamber. The distal part of fractured haptic was observed in the capsular bag. He underwent IOL exchange. The fracture site of the haptic was near the optic-haptic junction. This is the unique case report of a spontaneous fracture of an implanted posterior chamber polyimide IOL haptic, which implies the possibility of IOL haptic fracture in various haptic materials.

AB - A 57-year-old male patient visited our clinic for decreased visual acuity in the right eye for 10 days. He denied any trauma history, but recalled that the symptom developed after straining. He had undergone uncomplicated phacoemulsification and posterior chamber intraocular lens (IOL) implantation in the bag of the right eye 11 years ago. The IOL was a three-piece silicone polyimide-haptics design. On slit-lamp examination, the IOL optic and proximal part of nasal fractured haptic were found in the anterior chamber. The distal part of fractured haptic was observed in the capsular bag. He underwent IOL exchange. The fracture site of the haptic was near the optic-haptic junction. This is the unique case report of a spontaneous fracture of an implanted posterior chamber polyimide IOL haptic, which implies the possibility of IOL haptic fracture in various haptic materials.

UR - http://www.scopus.com/inward/record.url?scp=84898946360&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84898946360&partnerID=8YFLogxK

U2 - 10.4103/0301-4738.111195

DO - 10.4103/0301-4738.111195

M3 - Article

C2 - 23619497

AN - SCOPUS:84898946360

VL - 62

SP - 348

EP - 350

JO - Indian Journal of Ophthalmology

JF - Indian Journal of Ophthalmology

SN - 0301-4738

IS - 3

ER -